In this article, we design and evaluate a fractional-order proportional-integral-derivative (FOPID) controller for the regulation of total intravenous anesthesia (TIVA). In particular, the FOPID controller has been designed to regulate the patient’s depth of hypnosis (DoH), measured via the bispectral index (BIS) monitor, by coadministrating both the hypnotic and analgesic drugs used in TIVA. Separate tunings are obtained for the induction phase and for the maintenance phase. For the former, a methodology based on the minimization of the integrated absolute error (IAE) has been employed, while for the latter, the isodamping design approach has been applied. Experiments carried out on ten patients undergoing plastic surgery show that the FOPID controller meets the clinical requirements for each one of the ten patients without requiring any manual intervention from the anesthesiologist. Notably, the controller provided a fast rejection of disturbances without provoking harmful overdosing episodes.

In Vivo Performance Evaluation of an FOPID Controller for Closed-Loop Anesthesia

Paolino, Nicola;Schiavo, Michele;Latronico, Nicola;Paltenghi, Massimiliano;Visioli, Antonio
2025-01-01

Abstract

In this article, we design and evaluate a fractional-order proportional-integral-derivative (FOPID) controller for the regulation of total intravenous anesthesia (TIVA). In particular, the FOPID controller has been designed to regulate the patient’s depth of hypnosis (DoH), measured via the bispectral index (BIS) monitor, by coadministrating both the hypnotic and analgesic drugs used in TIVA. Separate tunings are obtained for the induction phase and for the maintenance phase. For the former, a methodology based on the minimization of the integrated absolute error (IAE) has been employed, while for the latter, the isodamping design approach has been applied. Experiments carried out on ten patients undergoing plastic surgery show that the FOPID controller meets the clinical requirements for each one of the ten patients without requiring any manual intervention from the anesthesiologist. Notably, the controller provided a fast rejection of disturbances without provoking harmful overdosing episodes.
File in questo prodotto:
File Dimensione Formato  
In_Vivo_Performance_Evaluation_of_an_FOPID_Controller_for_Closed-Loop_Anesthesia.pdf

solo utenti autorizzati

Licenza: DRM non definito
Dimensione 3.11 MB
Formato Adobe PDF
3.11 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11379/635875
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 0
social impact