Wheelchair ergometers are widely used in research, clinical practice, and sports environments. The majority of wheelchair ergometers are roller systems that allow for wheelchair propulsion in the personal wheelchair on one or two (instrumented) rollers. Oftentimes these systems are only statically calibrated. However, wheelchair propulsion is dynamic by nature, requiring a dynamic validation process. The aim of the current project was to present a low-cost portable system for the dynamic metrological verification of wheelchair roller ergometers, based on an instrumented reference wheel. The tangential force on the roller is determined, along with its uncertainty, from the reference wheel properties, and compared with the force measured by the ergometer. Uncertainty of this reference wheel system was found to be lower than the one of the ergometer used, indicating that this novel approach can be used for the metrological verification of ergometers.

Development of a Portable Low-Cost System for the Metrological Verification of Wheelchair Roller Ergometers

Lancini M.
Supervision
;
Spada P.
Formal Analysis
;
Muhametaj R.
Methodology
;
de Klerk R.
Investigation
;
Vegter R. J. K.
Formal Analysis
2023-01-01

Abstract

Wheelchair ergometers are widely used in research, clinical practice, and sports environments. The majority of wheelchair ergometers are roller systems that allow for wheelchair propulsion in the personal wheelchair on one or two (instrumented) rollers. Oftentimes these systems are only statically calibrated. However, wheelchair propulsion is dynamic by nature, requiring a dynamic validation process. The aim of the current project was to present a low-cost portable system for the dynamic metrological verification of wheelchair roller ergometers, based on an instrumented reference wheel. The tangential force on the roller is determined, along with its uncertainty, from the reference wheel properties, and compared with the force measured by the ergometer. Uncertainty of this reference wheel system was found to be lower than the one of the ergometer used, indicating that this novel approach can be used for the metrological verification of ergometers.
File in questo prodotto:
File Dimensione Formato  
bio_145_10_104501.pdf

accesso aperto

Descrizione: full text
Tipologia: Full Text
Licenza: PUBBLICO - Creative Commons 4.0
Dimensione 1.07 MB
Formato Adobe PDF
1.07 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11379/615050
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 0
social impact