Surgery for cochlear implant is a traumatic procedure, with inflammatory responses leading to immediate and delayed intracochlear changes, resulting in newly formed fibrous and bony tissue. This newly formed tissue is thought to affect speech perception with cochlear implants and can also play a role in causing device malfunctioning and soft failures. We present a case of left cochlear implant explantation and reimplantation in a 15-year-old girl, who experienced deterioration of speech perception and device failure associated with osteoneogenesis of the round window, which could represent a cause of cochlear implant failure. To avoid surgical trauma of the cochlear lateral wall, enlarged round window insertion rather than a cochleostomy, soft surgical techniques, and the application of steroids are all important issues to prevent new tissue formation, although special attention should also be given to the trauma of round window borders.

Osteoneogenesis at the Round Window: A Possible Cause of Cochlear Implant Failure?

Nassif, Nader
Membro del Collaboration Group
;
Redaelli de Zinis, Luca Oscar
Writing – Review & Editing
2023-01-01

Abstract

Surgery for cochlear implant is a traumatic procedure, with inflammatory responses leading to immediate and delayed intracochlear changes, resulting in newly formed fibrous and bony tissue. This newly formed tissue is thought to affect speech perception with cochlear implants and can also play a role in causing device malfunctioning and soft failures. We present a case of left cochlear implant explantation and reimplantation in a 15-year-old girl, who experienced deterioration of speech perception and device failure associated with osteoneogenesis of the round window, which could represent a cause of cochlear implant failure. To avoid surgical trauma of the cochlear lateral wall, enlarged round window insertion rather than a cochleostomy, soft surgical techniques, and the application of steroids are all important issues to prevent new tissue formation, although special attention should also be given to the trauma of round window borders.
2023
Altra università italiana
LS7_7 Surgery
Esperti anonimi
Inglese
Internazionale
ELETTRONICO
14
1
1
8
8
cochlear implant; device failure; osteoneogenesis; round window
no
Goal 3: Good health and well-being
3
info:eu-repo/semantics/article
262
Donati, Giulia; Nassif, Nader; Redaelli de Zinis, Luca Oscar
1 Contributo su Rivista::1.1 Articolo in rivista
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11379/613651
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 1
social impact