INTRODUCTION: Of the variables used by in vitro studies of resistance to sliding (RS) in orthodontics, sliding velocity (SV) of the wire is often the one farthest from its clinical counterpart. We investigated whether velocity influences the RS at values approximating the orthodontic movement. METHODS: A SS self-ligating bracket with a NiTi clip was fixed onto a custom-made model. Different shaped orthodontic SS wires of four sizes and two types (round, 0.020″ and 0.022″; rectangular, 0.016″×0.022″ and 0.017″×0.025″) were tested using an Instron® testing machine. Wires were pulled at four velocities (1×10-2 mm/s, 1×10-3 mm/s, 1×10-4 mm/s, 1×10-5 mm/s). Shapiro-Wilk test was used to evaluate the normal distribution of the data; two-way ANOVA was performed to compare means in the RS with wire characteristics and SV. Significance level was set at P<.05. RESULTS: RS was higher for rectangular wires, and for those with larger diameters. Lower SV was associated with lower RS, with wire type and size having an interaction effect. The RS relatively to SV can be represented as: RS ∝ α[ln(SV)]+β, where α and β are constants. CONCLUSIONS: At very low SV and low normal forces, SV influences the RS of SS archwires in orthodontic brackets, and the proportionality is logarithmic. Although respecting these parameters in vitro is challenging, quantitative evaluations of RS should be carried out at clinically relevant velocities if aiming at translational application in the clinical scenario.
In vitro evaluation of the influence of velocity on sliding resistance of stainless steel arch wires in a self-ligating orthodontic bracket
SAVOLDI, Fabio;VISCONTI, Luca;DALESSANDRI, Domenico;BONETTI, Stefano;PAGANELLI, Corrado
2017-01-01
Abstract
INTRODUCTION: Of the variables used by in vitro studies of resistance to sliding (RS) in orthodontics, sliding velocity (SV) of the wire is often the one farthest from its clinical counterpart. We investigated whether velocity influences the RS at values approximating the orthodontic movement. METHODS: A SS self-ligating bracket with a NiTi clip was fixed onto a custom-made model. Different shaped orthodontic SS wires of four sizes and two types (round, 0.020″ and 0.022″; rectangular, 0.016″×0.022″ and 0.017″×0.025″) were tested using an Instron® testing machine. Wires were pulled at four velocities (1×10-2 mm/s, 1×10-3 mm/s, 1×10-4 mm/s, 1×10-5 mm/s). Shapiro-Wilk test was used to evaluate the normal distribution of the data; two-way ANOVA was performed to compare means in the RS with wire characteristics and SV. Significance level was set at P<.05. RESULTS: RS was higher for rectangular wires, and for those with larger diameters. Lower SV was associated with lower RS, with wire type and size having an interaction effect. The RS relatively to SV can be represented as: RS ∝ α[ln(SV)]+β, where α and β are constants. CONCLUSIONS: At very low SV and low normal forces, SV influences the RS of SS archwires in orthodontic brackets, and the proportionality is logarithmic. Although respecting these parameters in vitro is challenging, quantitative evaluations of RS should be carried out at clinically relevant velocities if aiming at translational application in the clinical scenario.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.