A flow type quartz crystal microbalance (QCM) (bio)chemical sensor was developed for the real time determination of heavy metal ions that is suitable for environmental monitoring. A new process has been developed which enables to obtain surface-modified gold electrodes with high heavy metal ions complexing ability. The sensing performances of the piezoelectric sensor used in a flow-through setup were investigated by monitoring the frequency variation induced by the presence of heavy metal ions, such as copper and lead, as model ions, in aqueous media. X-Ray Reflectivity (XRR) and Atomic Force Microscopy (AFM) were carried out to characterize the unmodified and modified gold surfaces.
Polymer-coated quartz crystal microbalance chemical sensor for heavy cations in water
SARTORE, Luciana;PENCO, Maurizio;BERGESE, Paolo;BONTEMPI, Elza;DEPERO, Laura Eleonora
2009-01-01
Abstract
A flow type quartz crystal microbalance (QCM) (bio)chemical sensor was developed for the real time determination of heavy metal ions that is suitable for environmental monitoring. A new process has been developed which enables to obtain surface-modified gold electrodes with high heavy metal ions complexing ability. The sensing performances of the piezoelectric sensor used in a flow-through setup were investigated by monitoring the frequency variation induced by the presence of heavy metal ions, such as copper and lead, as model ions, in aqueous media. X-Ray Reflectivity (XRR) and Atomic Force Microscopy (AFM) were carried out to characterize the unmodified and modified gold surfaces.File | Dimensione | Formato | |
---|---|---|---|
66.pdf
gestori archivio
Tipologia:
Full Text
Licenza:
DRM non definito
Dimensione
2.65 MB
Formato
Adobe PDF
|
2.65 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.