The performance of the NucliSens easyMAG platform for the extraction of nucleic acid from different clinical specimens was compared with a manual procedure. A total of 308 specimens were analyzed: 209 plasma samples collected for virus detection and quantification of cytomegalovirus (CMV) and Epstein–Barr virus (EBV) (n = 70), and 29 for HIV genotyping for drug resistance. Linearity of extraction was tested on dilution series of CMVand EBV; the correlation coefficient (R2) for standard curves based on repeated extraction runs was 0.99 for CMV and EBV. Inter- and intrarun variability was in accordance with previous studies, and the correlation between automated and manual extraction was very high. The concordant results were 95.7% for CMV and 100% for EBV. The results of sequence analysis for HIV drug resistance showed a concordance in 24 of 29 specimens. The NucliSens easyMAG is extremely easy to perform, is automated, and resulted in a strong reduction of hands-on time compared with manual protocol.

Performance evaluation of the automated NucliSens EasyMAG nucleic acid extraction platform in comparison to QIAamp Mini Kit from clinical specimens.

PERANDIN, Francesca;BONFANTI, Carlo;MANCA, Nino
2009-01-01

Abstract

The performance of the NucliSens easyMAG platform for the extraction of nucleic acid from different clinical specimens was compared with a manual procedure. A total of 308 specimens were analyzed: 209 plasma samples collected for virus detection and quantification of cytomegalovirus (CMV) and Epstein–Barr virus (EBV) (n = 70), and 29 for HIV genotyping for drug resistance. Linearity of extraction was tested on dilution series of CMVand EBV; the correlation coefficient (R2) for standard curves based on repeated extraction runs was 0.99 for CMV and EBV. Inter- and intrarun variability was in accordance with previous studies, and the correlation between automated and manual extraction was very high. The concordant results were 95.7% for CMV and 100% for EBV. The results of sequence analysis for HIV drug resistance showed a concordance in 24 of 29 specimens. The NucliSens easyMAG is extremely easy to perform, is automated, and resulted in a strong reduction of hands-on time compared with manual protocol.
File in questo prodotto:
File Dimensione Formato  
easymag.pdf

gestori archivio

Tipologia: Full Text
Licenza: DRM non definito
Dimensione 575.03 kB
Formato Adobe PDF
575.03 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11379/19546
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 10
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 20
social impact