Thirteen failed stem of Total Hip Replacement were studied: 9 were Charnley THR from an homogeneous series, which gives an incidence of 2.4% of stem fractures with a follow-up of 9-16 years; 4 were Mueller THR. Fatigue fracture of the stem occurred by defective support of the proximal part of the femur, following resorption of the calcar. In all cases reactive tissue to foreign body particles, metal and polyethylene, was found where bone resorption occurred. In Mueller THR wear of the cup produced the large amount of polyethylene particles; in Charnley THR metal particles prevailed and corrosion of the stem is suggested to be the initiating factor.

Failure of the stem in total hip replacement. A study of aetiology and mechanism of failure in 13 cases.

PAZZAGLIA, Ugo;
1988-01-01

Abstract

Thirteen failed stem of Total Hip Replacement were studied: 9 were Charnley THR from an homogeneous series, which gives an incidence of 2.4% of stem fractures with a follow-up of 9-16 years; 4 were Mueller THR. Fatigue fracture of the stem occurred by defective support of the proximal part of the femur, following resorption of the calcar. In all cases reactive tissue to foreign body particles, metal and polyethylene, was found where bone resorption occurred. In Mueller THR wear of the cup produced the large amount of polyethylene particles; in Charnley THR metal particles prevailed and corrosion of the stem is suggested to be the initiating factor.
File in questo prodotto:
File Dimensione Formato  
Failure of stem.pdf

accesso aperto

Tipologia: Documento in Post-print
Licenza: PUBBLICO - Pubblico con Copyright
Dimensione 764.66 kB
Formato Adobe PDF
764.66 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11379/116358
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact